EFFECT OF WEAR OF THE ATOMIC-ABSORPTIONSPECTROMETER GRAPHITE FURNACEON THE ANALYTICAL ZONE TEMPERATURE
A. N. Kulik, A. N. Bugai, Yu. V. Rogul'skii,M. I. Zakharets, and L. F. Sukhodub UDC 543.42 One mechanism of increase in the heating temperature of the analytical zone of the graphite tube furnace of the atomic-absorption spectrometer as a result of wear is discussed. The influence of the furnace profile on its heating temperature is shown, and the calculated heating temperature of the standard tube is compared to the temperature in the special case where the cross-section area of the analytical zone of the furnace is reduced by half. Experimental proofs of a given temperature excess, as well as of a relation between the change in the tube profile and the decrease in its mass, are given. The dependence of temperature on the furnace mass and the number of measurements has been investigated. The necessity of taking into account the increase in the heating temperature of the furnace in the wear process has been considered. Institute of Applied Physics, NAS of Ukraine, 58 Petropavlovskaya Str., Sumy, 40030. Translated from Inzhenerno-Fizicheskii Zhurnal, Vol. 80, No. 1, pp. 149-155, January-February, 2007. Original article submitted February 14, 2005; revision submitted December 19, 2005.